There is a left femoral stent/graft seen on the left Hologic hip scan.

The accompanying radiograph shows the stent/graft.

Case Description:
There is a left femoral stent/graft seen on the left hip DXA. The accompanying radiograph also demonstrates the stent/graft. The stent/graft would be removed by the software from the soft tissue baseline. On a Hologic machine, the undo view may be checked to confirm that it is removed from the soft tissue baseline.
Credit:
Sarah L Morgan, MD, RD, CCD, The University of Alabama at Birmingham
References:
• Martineau, P., S. Bazarjani, and L.S. Zuckier, Artifacts and Incidental Findings Encountered on Dual-Energy X-Ray Absorptiometry: Atlas and Analysis. Semin Nucl Med, 2015. 45(5): p. 458-69.
• Bazzocchi, A., et al., Incidental findings with dual-energy X-ray absorptiometry: spectrum of possible diagnoses. Calcif Tissue Int, 2012. 91(2): p. 149-56.